– 1 –
TABLE OF CONTENTS
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.0 Create the calibration curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Define the fill quantity . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Prepare the samples . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Determination of reference moisture content . . . . . 4
2.4 Determination of the acual moisture content . . . . . . 4
2.5 Unlock the BM2 . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Change the calibration curve name . . . . . . . . . . . . . 5
2.7 Take over the values . . . . . . . . . . . . . . . . . . . . . . . . 6
2.8 Temperature compensation . . . . . . . . . . . . . . . . . . . 7
2.9 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.10 Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
– 8 –
With the following calculation the right TK1 can be calculated:
For example: Measure with your BM2 at 23°C a moisture content of 30%
and by the sample in the fridge (at app. 8°C) just 20%.
NOTE: Before measuring allow your BM2 to become conditioned to
the sample temperature.
2.9 Measuring range
In the menu item “Valid” is the validity of the calibration curve given. A
valid measurement is shown in the measuring window black, a invalid is
sown gray. This feature can be best explained by an example.
The goal is a valid range from 5% up to 25% moisture content because
for certain materials readings above 25% lose accuracy
Enter the following values in the shown order:
idx idx idx idx idx idx
(1) (2) (3) (4) (5) (6)
Idx(1) Idx(2) Idx(3) Idx(4) Idx(5) Idx(6)
50% 25.1% 25% 5% 4.9% 0%
00 1 1 0 0
The measruing range should e defined in the area of the measured
samples.
2.10 Completion
If all values of the created calibration curve are stored in the BM2
device, you can measure the material with the new calibration curve.
TK1=
30
=1.5
20
Vue de la page 0
1 2 3 4 5

Résumé du contenu

Page 1 - TABLE OF CONTENTS

– 1 –TABLE OF CONTENTS1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22.0 Create the calibration cur

Page 2

– 2 –1.0 INTRODUCTIONAselection of calibration curves are permanently strored in the BM2.However, if none of the provided calibration curves suit the

Page 3

– 3 –2.0 CREATE THE CALIBRATION CURVE2.1 Define the fill quantityThe fill quantity, that is, the amount of sample material placed in themeasuring ch

Page 4

– 4 –2.3 Determination of reference moisture contentCollect sample materialnot used in Samples 1 and 2. This will be used todeterminate the reference

Page 5 - CUSTOMER CALIBRATION

ELECTROMATIC E Q U I P M E N T C O., I N C.600 Oakland Ave., Cedarhurst, NY 11516–U.S.A.TEL: 516-295-4300 • FAX: 516-295-4399 CHECK•LINE®INSTRUMENTSB

Commentaires sur ces manuels

Pas de commentaire